Dr. Gregory House, PhD (Med. Phys.)
Computed Tomography (CT) revolutionized diagnostic medicine by providing three-dimensional cross-sectional imaging of the human body. By rotating a highly collimated X-ray tube and an opposing detector array around a patient, CT scanners measure projection data at multiple angles. Specialized mathematical reconstruction algorithms convert these raw attenuation projections into quantitative gray-scale maps of anatomical tissue density.
This article outlines the physics of CT imaging, Hounsfield Unit scaling, scan acquisition parameters, and standard clinical CT dose metrics.
Each pixel in a reconstructed CT image is assigned a value in Hounsfield Units (HU), also known as the CT Number. The HU is a normalized measurement of the linear attenuation coefficient (μ) of the tissue within that voxel relative to the attenuation coefficient of water (μ_w):
Because the attenuation coefficient varies with X-ray energy (beam spectrum), referencing the attenuation to water ensures that the CT numbers remain relatively stable across different scanner voltages (kVp).
By definition:
* Water: HU = 0
* Air: HU = -1000 (since μ_{air} ≈ 0)
* Fat: HU ≈ -120 to -80
* Soft Tissue (muscle, organs): HU ≈ +30 to +80
* Bone: HU ≈ +300 to +3000 (highly dense cortical bone)
In modern multidetector CT (MDCT) scanners, the patient couch moves continuously through the rotating gantry. This results in a helical (spiral) acquisition path.
The primary geometric parameter of helical scanning is the Pitch:
Where:
* I: The table travel distance in one 360^\circ rotation of the gantry.
* N: The number of detector rows.
* T: The width of a single detector row.
The projections collected by the detector are reconstructed using:
* Filtered Backprojection (FBP): A fast, analytical method that backprojects the filtered attenuation data onto a grid. It is computationally efficient but suffers from noise and streak artifacts in low-dose scans.
* Iterative Reconstruction (IR): Reconstructs images through an optimization loop that compares actual projections with simulated projections from a mathematical model. IR reduces noise and artifacts, allowing diagnostic images to be acquired at up to 50\% lower dose compared to FBP.
CT scans deliver higher radiation doses compared to projection radiography. To monitor and optimize patient safety, standardized dose indices have been defined:
If the pitch is < 1.0 (overlapping scans), the CTDI_{vol} increases.
DLP is measured in units of mGy·cm. The effective dose (E in mSv) can be estimated using tissue-specific conversion factors (k):
Computed Tomography physics requires a balance of geometric design, mathematical reconstruction, and dose optimization. By understanding Hounsfield scaling, optimizing scan pitch, and monitoring dose indices like CTDI_{vol} and DLP, medical physicists ensure that diagnostic scans are performed with high clinical efficacy at minimal radiation hazard, adhering to the core tenets of radiation safety.